40
BASIC OPERATIONS
Characteristics of each examination course
The details automatically set in each examination course is as shown below.
5 Press and display the measurement screen.
Selection by examination/
purpose
Characteristics
Screening
Measures basic information related to corneas and entire ocular
refraction.
Pupillometry
Measures pupil diameters, REF values, and high-order aberra-
tions in dark and bright places.
Dry eyes
Measures ten times successively in an interval of one second for
about ten seconds. This indicates progressive change in the
ocular characteristics.
Cataract
Provides useful information for the best Hartmann selection
before operation as well as used to check the condition after the
operation. It has a longer exposure time to make it easier to
measure the point image with the cataract eye.
Over refraction Checks the condition when glasses are worn.
Custom
Provides measurement and report output according to the set-
ting made in the Setup screen.
Selection by exam-
ination/purpose
Measure-
ment target
No. of mea-
surement
Hartmann
exposure time
Fixation
target
Report
format
Screening
Cornea/
refraction
3 times for
each eye
5: Normal
(256x64)
Level 3 Both eyes
Pupillometry
Cornea/
refraction
3 times for pho-
topic/3 times
for scotopic
5: Normal
(256x64)
Automatic Both eyes
Dry eyes
Cornea/
refraction
10 times for
each eye
5: Normal
(256x64)
Level 3
One eye at
each time
Cataract
Cornea/
(refraction)
3 times for
each eye
ON (300ms) Level 2 Both eyes
Over refraction Ref only
3 times for
each eye
5: Normal
(256x64)
Level 3 Both eyes
The setting of each measurement course can be changed. Refer to page 175 for
the detail.
In case of the Selection of examination interface, tap the desired measurement
course, and the measurement screen is brought up immediately.
When the check button is checked, tap the desired measurement course, and the
measurement screen is brought up immediately.
OK button