SETUP OF SELECTION OF EXAMINATION MEASUREMENT
SCREENING SETTINGS
This is to set the Screening measurement course.
1 Tap "Screening" in the index to select.
NUMBER OF MEASUREMENT SETTING
1 Tap the "Number of measurement" in the items.
2 Tap the to set the number of measurement.
3 Press to complete the setting.
SETTING HARTMANN EXPOSURE TIME
Hartmann exposure time can be selected. Prolonging the exposure time can reduce the mea-
surement error that occurs in the hard-to-measure eye. The default setting is 100ms(Normal).
1 Tap the of "Hartmann Exposure Time" displayed in the item "Tem-
porary Setting".
2 Tap the and select "100ms (Normal)", "300ms", "400ms" or
"500ms" by tapping.
3 Setting is complete.
"100ms(Normal)": The Hartmann exposure time is set to 100ms. (Standard setting)
"300ms" : The Hartmann exposure time is set to 300ms.
"400ms" : The Hartmann exposure time is set to 400ms.
"500ms" : The Hartmann exposure time is set to 500ms.