114
FUNCTION SETTING USING SETTINGS SCREEN
LIST OF SETUP ITEMS
Setup items are formed with eight indexes for each subject.
"MEASURE" Set items related to measurement functions
"INITIAL" Set items related to the initial status after power on
"DISPLAY" Set items related to screen display
"PRINT INT" Set items related to output from the internal printer
"PRINT EXT" Set items related to output from the external printer
"COMM" Set items related to data input/output between the external device
"NETWORK" Set items related to input/output using the LAN
"SPECIAL" Set items related to maintenance (for service engineer only)
MEASURE (TEMPORARY SETTING)
MEASURE contains settings related to temporarily changeable measurement functions.
Unlike initial setting (INITIAL), this function is used to temporarily perform measurements
under certain conditions.
After power on, clearing all measurement values, print-out and data output, the system
returns to the initial status.
Set item display Setting Change button Set contents Initial value Page
R-L Measure
BINO Set to both-eye measurement
BINO 124
MONO Set to single-eye measurement
Auto Mode
FULL AUTO Set to (FULL) AUTO mode
FULL AUTO
125AUTO TRACKING Set to AUTO TRACKING mode
MANUAL Set to MANUAL mode
Cont. Cycle
1
Sets the number of continuous measurements under
AUTO mode, using up/down buttons.
3 125
2
3
4
5
6
7
8
9
10
Hartmann
exposure time
100ms(Normal)
The Hartmann exposure time is set to 100ms.
(standard setting)
100ms 126
300ms The Hartmann exposure time is set to 300ms.
400ms The Hartmann exposure time is set to 400ms.
500ms The Hartmann exposure time is set to 500ms.
Target Brightness
Level 1 Target brightness is set to the darkest of 4 levels.
Level 3 126
Level 2 Target brightness is set to the 2nd darkest of 4 levels.
Level 3 Target brightness is set to the 3rd darkest of 4 levels.
Level 4 Target brightness is set to the brightest of 4 levels.
Cyl sign
+ Display style of astigma sign is set to "+".
- 127- Display style of astigma sign is set to "-".
MIX Display style of astigma sign is set to "MIX".
Setting VD
0mm VD value is set to 0mm (contact lens).
12nm 12712mm VD value is set to 12mm (eyeglass lens).
13.75mm VD value is set to 13.75mm (eyeglass lens).

Table of Contents