113
MAINTENANCE
MAINTENANCE
MAINTENANCE CHECKUPS
Maintenance of this instrument includes maintenance items by the user and maintenance items by
the manufacturer.
USER MAINTENANCE ITEMS
MANUFACTURER MAINTENANCE ITEMS
CAUTION
To prevent damage and injury, do not install the instrument on an
uneven, unsteady or sloped surface.
Item Inspection time Contents
Inspection Before using
• The instrument works properly.
• The measuring window must be free of stains
and/or flaws.
• Confirm whether the foreign object is attached
to the measuring nozzle and the area around
the measuring nozzle.
• Air check
• Confirm that the safety stopper setting and
measuring nozzle do not move to the patient's
side
beyond the safety stopper setting position.
Cleaning When the part is stained
• Measuring window
• Instrument cover, control panel, etc.
Replacement As required • Printer paper
Item Checking time Contents
Cleaning each
component
Within 12 months
• Cleaning outer covers
• Cleaning the optical system
• Cleaning Power supply unit
Operation check Within 12 months
• Checking the main body operation
• Checking for looseness of the chinrest unit
Accuracy check Within 12 months
• Confirming the intraocular pressure measure-
ment functions (using special tools)
• Confirming the cornea thickness measurement
functions (using special tools)