137
GENERAL INFORMATION ON USAGE AND MAINTENANCE
ELECTROMAGNETIC COMPATIBILITY
This product conforms to the EMC standard IEC 60601-1-2:2014+AMD1:2020 (Ed.4.1).
The expected electromagnetic environment for the whole life cycle is home medical treatment environ-
ment.
a)MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the ACCOMPANYING
DOCUMENTS.
b)Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
c)The use of ACCESSORIES, transducers and cables other than those specified, with the exception
of transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replace
-
ment parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY
of the EQUIPMENT or SYSTEM.
d)The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be observed to verify
normal operation in the configuration in which it will be used.
e)The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other than those
specified may result in increased EMISSION or decreased IMMUNITY of the EQUIPMENT or SYSTEM.
f) Do not use the devices generating electromagnetic waves within 30cm from all the parts of the
instrument and system. Those devices may have influence on this instrument.
Item Length (m) Shield Ferrite core
AC power cord (AC 100 V/120 V) 1.5 No No
AC power cord (AC 230 V/240 V) 3.0 No No
Power cord (for computer) 1.5 No Yes
Power cord (for AC adapter) 1.0 No No
LAN cable 8.0 Yes Yes
Barcode reader cable 1.5 Yes Yes
Personal computer ― ― ―
USB barcode reader ― ― ―
AC adapter ― ― ―
Guidance and manufacturer's declaration - electromagnetic emissions
The TRK-3 OMNIA is intended for use in the electromagnetic environment specified below.
The customer or the user of the TRK-3 OMNIA should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The TRK-3 OMNIA uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause any inter-
ference in nearby electronic equipment.
RF emissions
IEC 61000-3-2
Class B
The TRK-3 OMNIA is suitable for use in all establishments including
domestic and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies