EasyManua.ls Logo

Topcon TRK-3 OMNIA - Specifications and Performance

Topcon TRK-3 OMNIA
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
129
SPECIFICATIONS AND PERFORMANCE
SPECIFICATIONS AND PERFORMANCE
Conformity standards
ISO 8612: 2009 Ophthalmic instruments–Tonometers
ISO 10342: 2010 Ophthalmic instruments–Eye refractometers
ISO 10343: 2014 Ophthalmic instruments–Ophthalmometers: Type B
REF measurement
Measuring range Spherical refractive power:
25 D to +22 D
(Display value resolution: 0.12 D or 0.25 D)
Astigmatic refractive power: 0D to ±10 D
(
Display value resolution:
0.12 D or 0.25 D)
Direction of astigmatic axis: 0° to 180° (
Display value resolution:
1° or 5°)
(where, spherical refractive power + astigmatic refractive power +22 D,
or spherical refractive power + astigmatic refractive power ‒25 D)
The dioptric powers are indicated with reference wavelength λ
d
= 587.56
nm.
Measurable mini-
mum pupil diameter
Φ2.0 mm
PD measuring
range
20 mm to 85 mm (Display value resolution: 1 mm)
KRT measurement
Measuring range Cornea curvature radius: 5.00 mm to 10.00 mm
(Display value resolution: 0.01 mm)
Corneal refractive power: 67.50 D to 33.75 D
(Display value resolution: 0.12 D or 0.25 D)
(where, corneal refractive power=1.3375)
Direction of corneal principal meridian: 0° to 180°
(Display value resolution: 1° or 5°)
Corneal diameter measurement
Measuring range 2.00 mm to 14.00 mm (Display value resolution: 0.25 mm)
Intraocular pressure measurement
Measuring range 7 mmHg to 60 mmHg (Display value resolution: 1 mmHg)
Average value resolution: 1 mmHg or 0.1 mmHg
Measuring mode 7 to 30 mmHg or 30 to 60 mmHg, selectable range
Corneal thickness measurement
Measuring range 0.400 mm to 0.730 mm (Display value resolution: 0.001 mm)