64
BASIC OPERATIONS
BASIC OPERATIONS
MEASUREMENT IN REF/KRTTONO/PACHY CONTINUOUS MEASUREMENT MODE
DISPLAYING MEASUREMENT VALUES
With regard to measurement values, for REF, KRT, TONO and PACHY, data of the latest measure
-
ment (only for TONO/PACHY, latest 3 measurements) are displayed on the control panel.
numerical value only: Measurement was done correctly.
numerical value with [ ]: When the reliability of measurement is low. (only TONO)
ERROR: Measurement was not done correctly.
OVER: When IOP value exceeds 60 mmHg.
NOTE
In TONO average value display, low-reliability numerical data with [ ] are
not added to average value calculation.
However, if all measurement data are numerical data with [ ], average
value calculation is done using these data.
For explanation of the messages on the control panel screen, refer to
MESSAGE LIST” on page 120.
When data is printed out, the values measured using the Start button or
the control lever’s MEASUREMENT switch are displayed with an M mark
next to them. (TONO/PACHY only)
When the “Auto Print” is set to the “OFF” in the “Printer(Common)1” on the
Setup screen, print out measurement results by tapping the
button, as necessary.
Print out