63
BASIC OPERATIONS
BASIC OPERATIONS
MEASUREMENT IN REF/KRTTONO/PACHY CONTINUOUS MEASUREMENT MODE
2 Alignment starts automatically. While moving the main body toward the patient, the focus of
Measurement screen is changed, then measurement is performed.
The measuring head moves automatically to the other eye side and measurement is per-
formed. The measurement results are displayed.
NOTE
• When the “Full Auto” of “R/L mode” is selected in the “Function(T/P)1” on
the Setup screen, the instrument measuring head moves automatically to
the other eye side for measuring. If the patient closes or blinks their eyes
at the time of the right-and-left eye change, the change may be unable to
be performed correctly. If the “Manual” of “
R/L mode” is selected, press the
button or button of the other eye side. If the “Auto (RL)” of “R/L
mode” is selected, the measuring head moves automatically to the other
eye side, however a measurement is not performed.
• Auto Print (available only under Auto mode)
When the “Auto Print” is set to “ON” in the “Printer(Common)1” on the
Setup screen, measurement results are printed out automatically after
measuring the right and left eyes. (See page
108.)
• When “Tap measurement position to measure after manual focus.” is dis-
played, please confirm if the patient's eye fits normal conditions for mea-
suring. Then tap the pupil on the control panel again.
• When the alignment status has continued for more than 30 seconds,
“Alignment error occurred. Focus manually and tap pupil on screen to
restart measurement. (E03003)” is displayed and the alignment stops.
• To stop alignment in the middle, tap the button. It is possi-
ble to stop alignment also by tapping the control panel anywhere, while an
button is displayed. When “Alignment stopped. Tap pupil
on screen to continue measurement.” is displayed, please tap the pupil on
the control panel again.
• When PACHY measurement is performed, AUTO/MANUAL display is
changed to “PACHY measurement in progress.”. When TONO measure
-
ment is performed, the mark is changed to “TONO measurement in prog-
ress.”. The mark indicates the current status of measurement.
• If the alignment status has continued for more than 5 seconds in PACHY
measurement, the measurement stops to change to TONO measurement.
• If the start button is tapped before all PACHY measurements are com-
plete, PACHY data is not measured for remaining number of times is
treated as error data.
R L
Alignment stop
Alignment stop