106
SETTING FUNCTIONS ON SETUP SCREEN
Function(R/K)1
Function(R/K)2
Function(REF)
Function(KRT)
Descriptions Options Details
Initial value
(Region)
A C E D J G
Cont. Cycle
1-10
Set by number display
The number of continuous measurements. 3
Add Measure
0-99
Set by number display.
When the measurement is error,
set the number of times of remeasurement.
1
Continuous fog
Every time Continuous fog is applied every time.
Once Continuous fog is applied only once before the 1st measurement.
R/L mode
Manual
The other eye button is used to manually switch between right
and left eyes in continuous measurement mode.
Full Auto
Right and left eye switching is automatically performed in continu-
ous measurement mode before measurement.
Auto (RL)
Right and left eye switching is automatically performed in continu
-
ous measurement mode, but no measurement is performed.
Touch Measure
OFF Touch measurement is not performed in Manual mode.
ON Touch measurement is performed in Manual mode.
Sph/Cyl step
0.12 Sph/Cyl is displayed by 0.12D step.
0.25 Sph/Cyl is displayed by 0.25D step.
Axis step
1 Axial angle is displayed by 1° step.
5 Axial angle is displayed by 5° step.
Cylinder sign
Cylinder sign is “‒”.
+ Cylinder sign is “+”.
MIX Cylinder sign is “‒” and “+”.
Descriptions Options Details
Initial value
(Region)
A C E D J G
Model Eye Measure Mode Execute
Clears the measurement value and transitions to measurement
standby for model eye measurement.
Eyelid function
OFF Disables the eyelid function.
ON Enables the eyelid function.
Eyelid Force Meas. time
0-99
Set by number display.
Set the time [Second] from when the eyelid detection message
is displayed until the forced measurement starts.
99
Descriptions Options Details
Initial value
(Region)
A C E D J G
VD
0.00 VD value is set to 0mm (contact lens).
12.00 VD value is set to 12.00mm (eyeglass lens).
13.75 VD value is set to 13.75mm (eyeglass lens).
REF average
OFF REF average is not displayed.
ON REF average is displayed.
Descriptions Options Details
Initial value
(Region)
A C E D J G
D or mm
D D (diopter) of corneal refractive power.
mm mm of corneal curvature.
HV or R1R2
HV
Corneal curvature radius measurement result on screen
is displayed by HV.
R1R2
Corneal curvature radius measurement result on screen
is displayed by R1R2 (flat/steep meridian).
KRT Display Unit
OFF KRT unit is not shown.
ON KRT unit is shown.