GENERAL SAFETY INFORMATION
CONTRAINDICATION
Ensuring the Safety of Patients and Operators
To prevent corneal damage, do not measure a patient with corneal disease or one who's had corneal surgery.
WARNING
Ensuring the Safety of Patients and Operators
Be careful not to hit the patient's eyes or nose with the instrument during operation.
The patient may be injured.
Preventing Electric Shocks and Fires
To avoid fire and electric shock, install the instrument in a dry place free of water and other liquids.
To avoid fire and electric shock, do not put cups or other containers with liquids near the instrument.
To avoid fire in use, do not use the power inlet, power connector, or power plug with dust or liquid adhering to or
accumulating on them.
To avoid fire in the event of an instrument malfunction, immediately turn OFF the POWER switch ā ā and discon-
nect the power plug from the outlet if you see smoke coming from the instrument, etc. Do not install the instrument
where it is difficult to disconnect the power plug from the outlet. Ask your dealer for service.
To avoid fire and electric shock, do not open the cover. Ask your dealer for service.
Modification of this instrument is not permitted.
If the product is used for a long period of time with the power plug plugged into the outlet, dust may collect
between the outlet and the power plug, causing electric leakage due to the adhesion of water and causing a fire.
Clean the space between the outlet and the power plug regularly.
(There is a risk of electric leakage or fire due to the accumulation of dust and the adhesion of moisture.)
To avoid injury from electric shocks, do not connect anything other than the specified external equipment to the
external I/O terminals.
Be sure to connect the power plug to an AC 3-pin receptacle equipped with grounding. Connection with recepta-
cle without grounding may cause fire and electric shock in case of short-circuiting.
Connect only items that have been specified as part of the ME system or that have been specified as being com-
patible with the ME system.