TROUBLESHOOTING
TROUBLESHOOTING
MESSAGE LIST
Initializing. Please wait... Displayed when the main unit is being initialized at startup.
Printer device might not be under
normal operations after initializa
-
tion.
Tap OK if using without printer.
Else, please reboot off/on sys-
tem.
(E01003)
An error has occurred with the built-in printer.
The printer may not be working properly, but you can continue to use it.
Even if you want to continue using the instrument, please contact your service
personnel as soon as possible.
Failure to start network connec-
tion.
Check network cable and con-
nection settings.
(E01004)
Displayed when the initialization of the network connection fails. Check if the
LAN cable connection and network settings are correct.
System detected to not be under
normal operations after initializa-
tion.
Please reboot off/on system.
(E01005)
Displayed when there is a problem with the internal storage.
Please turn the power of the main unit off and then on again.
If this message is displayed repeatedly even after turning the power off and
then on again, contact your service personnel.
System detected to not be under
normal operations after initializa-
tion.
Please reboot off/on system.
(E01101)(E01102)(E01103)
Displayed when the connection between the auto-alignment sensor and this
instrument is abnormal or not connected. Please turn the power of the main
unit off and then on again. If this message is displayed repeatedly even after
turning the power off and then on again, contact your service personnel.
Chin rest failed to work properly.
Please reboot off/on system.
(E01104)
Displayed when the connection between the chinrest and this instrument is
abnormal or not connected.
Please turn the power of the main unit off and then on again. If this message is
displayed repeatedly even after turning the power off and then on again, con
-
tact your service personnel.
For the safety of the patient,
please be sure to set the safety
stopper before measuring.
Will this message be displayed at
startup from the next time?
Displayed to prompt the setting of the safety stopper. Set the safety stopper.
System detected to not be under
normal operations after initializa-
tion.
Please reboot off/on system.
(E01202)
Displayed when the connection between the target fixation sensor and the
device is abnormal or not connected. Please turn the power of the main unit off
and then on again. If this message is displayed repeatedly even after turning
the power off and then on again, contact your service personnel.
Checking air pressure used for
TONO measurement.
Air will come out of nozzle three
times.
Avoid putting face on chin rest
until complete.
Check the air pressure used for TONO measurement.
Air comes out of the measuring nozzle three times.
Please do not put face on chinrest until complete.
Return measuring head to central
position and turn off power?
Displayed to prompt confirmation whether to move the chinrest and the mea-
surement head to their last positions.
End of operation in progress.
Please wait until complete.
Displayed that the end operation is in progress.
End of operation complete.
Please turn power off on unit.
Displayed when the end operation is completed.
Measurement complete. Displayed when normal measurement is completed for the set number of
times.