GENERAL INFORMATION ON USAGE AND MAINTENANCE
PATIENT ENVIRONMENT
When the patient or inspector comes into contact with the devices (including the connecting devices) or
when the patient or inspector is in contact with the person that touches the devices (including the con
-
necting devices), the patient environment is shown below.
In the patient environment, use devices conforming to IEC 60601-1. If you are compelled to use any
device not conforming to IEC 60601-1, use an insulation transformer.
Note 1: Use the personal computer conforming to IEC 62368-1.
Note 2: Do not remove the cover from the personal computer.
Note 3: Use the insulation transformer conforming to IEC 60601-1.
WARNING
Connect only items that have been specified as part of the ME
system or that have been specified as being compatible with the
ME system.
CAUTION
• Do not connect an additional power strip or an extension cord to
the system.
• The total 1kVA is the maximum allowable load of the auxiliary
power supply socket for the insulation transformer, which is pro
-
vided for the system.
Do not connect the device exceeding this capacity.
• Use the auxiliary power supply socket of the insulation trans-
former to power only a device that will be a component of the
system.
• It is dangerous to connect any device which is not used as a
component of the system, to the insulation transformer.
• When the insulation transformer is not used, the personal com-
puter and the monitor for the personal computer must be
installed out of the patient environment.
Do not use multiple outlet strip in the
patient environment.
Be sure to connect the power supply
to commercial power.
Radius of 1.5 m
2.5m
1.5m1.5m
• Personal computer
(Note 1) (Note 2)
• Monitor display for PC
(Note 1)
• Insulation transformer
(Note 3)
• Mouse
Suitable device to use in the patient
environment