PREPARATION BEFORE MEASUREMENT
AIR CHECK
Before TONO/PACHY measurement, an air check is required.
This instrument is equipped with a function for checking correct operations of the measurement sys-
tem inside the instrument.
1 On the measurement standby screen of Auto mode in T/P mode, tap the button.
2 Then, the confirming message of Air check/End operation is displayed.
3 Press the button. Then, air comes out from the measuring nozzle automatically.
4 Make sure the message box of “Pressure used for tonometry OK.” is displayed on the control
panel.
5 Press the button, return to Measurement screen.
Air check
OK
OK