46
BASIC OPERATIONS
BASIC OPERATIONS
PREPARATION BEFORE MEASUREMENT
NOTE
If “Pressure detected out of standard performance. Check and clean
measuring nozzle before trying again. Refer to manual for cleaning
instructions. (E09104)” is displayed, the condition is not normal. Check if
there are any obstacles in front of the measuring nozzle.
If there is any, remove it, press the button and do the check again.
If there is no object, a failure is suspected. Turn the POWER switch to OFF,
unplug the power cord, and call your dealer or TOPCON at the address
printed on the back cover of this manual.