GENERAL INFORMATION ON USAGE AND MAINTENANCE
GENERAL INFORMATION ON USAGE AND MAINTENANCE
INTENDED PATIENT POPULATION
The patient who undergoes an examination by this instrument must maintain concentration for a few
minutes and keep to the following instructions:
• To fix the face to the chinrest, forehead rest.
• To keep the eye open.
• To understand and follow instructions when undergoing an examination.
However, patients who are able to perform the above actions but have any of the product's contraindi-
cations are excluded.
INTENDED USER PROFILE
Ophthalmologists, optometrists, orthoptists (ORT), Ophthalmic Medical Assistant (OMA), other certified
healthcare professionals*, or persons who received training under the guidance of a doctor.
* Other certified healthcare professionals; Qualified person authorized to use this product by local
laws and regulations.
ENVIRONMENTAL CONDITIONS OF USE
Temperature: 10°C to 35°C
Humidity: 30% to 90% (without condensation)
Pressure: 800 hPa to 1060 hPa
STORAGE, USAGE PERIOD
1. Environmental conditions
Temperature: −10°C to 55°C
Humidity: 10% to 95% (without condensation)
Pressure: 700 hPa to 1060 hPa
2. When storing the instrument, ensure that the following conditions are met:
(1) The instrument must not be splashed with water.
(2) Store the instrument away from environments where pressure, temperature, humidity,
ventilation, sunlight, dust, salty/sulfurous air, etc. could cause damage.
(3) Do not store or transport the instrument on a slanted or uneven surface or in an area
where it is subject to vibrations or instability.
(4) Do not store the instrument where chemicals are stored or gas is generated.
3. Normal life span of the instrument:
8 years from delivery providing regular maintenance is performed [TOPCON data]
Follow the respective handling methods for the transportation and storage conditions of external I/O
devices.
TRANSPORT CONDITIONS
Temperature: −40°C to 70°C
Humidity: 10% to 95%
Pressure: 700 hPa to 1060 hPa