121
TROUBLESHOOTING
Start measurement by tapping
pupil on screen.
Displayed when “Full Auto” measurement is possible. Tap the pupil to start
measurement automatically.
Ready to measure. Displayed when measurement can be performed in Manual mode.
Tap measurement position to
measure after manual focus.
Displayed as instruction of the measurement procedure in manual mode. Tap
the measuring head forward/backward button to adjust the measuring head
position, and then tap the measurement point.
Press button on control lever to
measure, after manual adjust-
ment.
When “Touch measure” in “Function(R/K)1” on the Setup screen (3/9) and
“Function(T/P)1” on the Setup screen (4/9) is “OFF”, it is displayed as instruc-
tion of the measurement procedure in manual mode. Tap the measuring head
forward/backward button to adjust the measuring head position, and then tap
the start button.
Auto alignment is progress.
Please wait a moment.
Displayed while auto-alignment is performed in R/K and T/P measurement
modes.
Alignment error occurred.
Focus manually and tap pupil on
screen to restart measurement.
(E03003)
Displayed when the XYZ positions are not aligned. Tap the measuring head
forward/ backward button to adjust the measuring head position, and then tap
the pupil again to align the pupil with the center of the screen.
Measurement interrupted due to
covering eyelids of patient.
Try again with patient's eyes wide
open.
(E03004)
Displayed when the patient's eyelids cover the pupil. Tell the patient to open
their eyes as wide as possible.
Alignment stopped.
Tap pupil on screen to continue
measurement.
Displayed when Alignment stop button is tapped.
Reaching set limit of stopper.
Move measuring head.
Displayed when the measuring head position has reached the safety stopper.
Check eyelid. Displayed when the subject's eyelid is over the pupil. Instruct the subject to
keep the eye open wider.
Focus error occurred.
Focus forward/backward manu-
ally and
tap pupil on screen to restart
measurement.
(E03008)
Displayed when the Z position cannot be detected in T/P measurement mode.
Tap the measuring head forward/backward button to adjust the measuring
head position, and then tap the pupil again on the screen to bring it to the cen
-
ter of the screen.
NEAR LIMIT Displayed when the measuring head is too close to the patient's eye.
Move the instrument as far from the patient's eye.
TOO CLOSE Displayed when the measuring head is too close to the patient's eye.
Move the instrument as far from the patient's eye.
FAR LIMIT Displayed when the measuring head is too far from the patient's eye.
Move the instrument as close to the patient's eye.
Measurement mode cannot be
established. Select one or more
of REF / KRT / TONO / PACHY.
(E05001)
None of the measurement modes, REF / KRT / TONO / PACHY, is selected.
Select one or more of REF / KRT / TONO / PACHY.
KRT measurement in progress. Displayed while measuring the cornea curvature radius.
REF measurement in progress. Displayed while measuring the spherical refractive power.
PACHY measurement in prog-
ress.
Displayed while measuring cornea thickness measurement.
TONO measurement in progress. Displayed while measuring the intraocular pressure measurement.
Please tap the camera icon. Displayed when the image for corneal diameter measurement can be taken
with alignment.
Measurement stopped.
Tap pupil on screen to continue.
Displayed when measurement is stopped during measurement.
If you want to continue the measurement, tap the pupil again.