TROUBLESHOOTING
Start alignment by tapping pupil
on screen.
Displayed on the Capture screen for corneal diameter measurement. Tap the
pupil on the screen if you need alignment.
Change of measurement head
position in progress (R/L).
Displayed while switching the measurement target eye of the measuring head.
Safety stopper set out of range.
Change to within 0 mm to 30 mm.
(E09001)
Displayed when the Z-axis position when the OK / Apply button is pressed is
outside the range of 0 mm to 30 mm on the safety stopper screen. Please set
within the range.
Check air pressure used for
TONO measurement?
NOTE: Air comes out of nozzle.
Displayed for confirmation of whether to perform an air check.
Air pressure used for TONO
measurement OK.
Displayed when the normal operation of the air check is confirmed. Press the
OK button.
Air pressure detected out of stan-
dard performance.
Check and clean measuring noz-
zle before trying again.
Refer to manual for cleaning
instructions.
(E09103)(E09104)
Displayed when an anomaly occurred during air check. Check the measuring
nozzle for any foreign matter.
MANDATORY field missing:
Patient ID.
(E09201)
Displayed when “Required Patient ID” of “System 1” on the Setup screen (1/9)
is set to “ON” and trying to output without inputting the patient ID. Enter the
patient ID before outputting.
MANDATORY field missing: Phy-
sician ID.
(E09202)
Displayed when “Physician ID request” of “Physician ID” on the Setup screen
(7/9) is set to “ON” and trying to output without inputting the physician ID. Enter
the physician ID before outputting.
Printer paper end. Refill if neces-
sary and try again.
Printer paper is used up. Supply printer paper.
Close the printer cover before the
measurement.
The printer cover is open. Close the cover until it clicks.
Data output via the net-
work...please wait for a while.
Displayed while outputting data via the network.
Data output via network com-
plete.
Displayed when the output of data via the network is completed.
Confirm setting “Preset(All)”?
Note:
This sets printing out all mea-
sured results.
Set the print settings of the built-in printer to All format.
Confirm setting to “Pres-
ent(Avg)”?
Note:
This prints out only average mea-
suring results.
Set the print settings of the built-in printer to Avg format.
Confirm setting “Preset (Clas-
sic)”?
Note:
This changes print out of the
measuring result form of equiva-
lent to conventional RM KR-8900
Classic2.
Set the print settings of the built-in printer to Classic format.
No data to output. Displayed when there is no data to be output while attempting manual printing
with the printer set to OFF.