TROUBLESHOOTING
Output data settings not set.
Confirm settings.
Displayed when all output settings are OFF. Confirm that the output setting is
in the correct way.
Printer failure.
Turn off/on unit and try again.
(E10010)
Displayed when the printer unit does not move normally, such as when the cut
-
ter does not move, so call your service personnel.
Error occurred during printer out-
put.
Check that the printer cover is not
open.
After checking, output the data
again.
(E10011)
Displayed when an error occurs during printer output.
Check to see if the printer cover is open and try outputting again.
Error occurred during printer out-
put.
The printer ran out of paper.
Please refill the paper and output
again.
(E10012)
Displayed when an error occurs during printer output.
Refill the printer paper and output again.
Please wait until the next mea-
surement is ready.
Displayed during initialization of measurement data and measurement environ-
ment.
Please wait.
Change in settings detected and
are not saved.
Confirm to discard settings to exit
screen.
Not saved after changing the settings.
To cancel the setting and exit from the Setting screen, press the OK button.
There is data being edited.
Do you want to discard it?
After adjusting the alignment bar in the Corneal Diameter Measurement
screen, the Measure button was not pressed, or after adjusting the safety stop
-
per position in the Safety Stopper screen, the Apply button was not pressed. If
you want to discard the adjusted value, press the OK button.
Applying network settings.
Please wait for a moment.
Displayed when the connection settings are being applied due to a change
made to the network settings.
Incorrect setting value (1-223 for
first octet).
(E12002)
Displayed when the first octet of “IP address”, “Default gateway”, “Primary
DNS server” or “Secondary DNS server” of “Network 3” on the Setup screen
(1/9), and “IP address” of “DICOM MWL Server” on the Setup screen (8/9) is
set to a value out of the specified input range. Enter a value within the input
range.
Incorrect setting value.
Enter value between 0 and 255
for IP address.
(E12003)
Displayed when any octet other than the first octet in “IP address”, “Default
gateway”, “Primary DNS server” or “Secondary DNS server” of “Network 3” on
the Setup screen (1/9), and “IP address” of “DICOM MWL Server” on the
Setup screen (8/9) is set to a value out of the specified input range. Enter a
value within the input range.
Incorrect setting value.
Enter value between 0 and 255
for IP address.
(E12004)
Displayed when the “Subnet mask” of “Network 3” on the Setup screen (1/9) is
set to a value that does not conform to the input rule. Enter a value that con
-
forms to the “Subnet mask” input rule.
Provide valid number value NNN
to NNN.
Displayed when an attempt has been made to set a value outside of the speci-
fied setting range. Enter a value within the specified input range.
“Fixed Physician ID” setting is
now turned on.
If you want to change the Physi-
cian ID, turn off “Fixed Physician
ID”.
Displayed when “Fixed Physician ID” of “Physician ID” on the Setup screen (7/
8) is set to ON and trying to enter the physician ID. Set the Fixed Physician ID
to OFF if you need to enter the physician ID.
Connection error(E14101) Displayed in case of a network error other than errors E14102 through E14112
and E14115. Call your service personnel.