107
SETTING FUNCTIONS ON SETUP SCREEN
Function(T/P)1
Function(T/P)2
Function(TONO)
Descriptions Options Details
Initial value
(Region)
A C E D J G
Meas. Start: Measure Mode
TONO
TONO is selected for the measurement mode at the start of T/P
measurement when “
Meas. Start: Keep Prev. Mode” is INIT.
T/P
TONO/PACHY is selected for the measurement mode at the start
of T/P measurement when “Meas. Start: Keep Prev. Mode” is
INIT.
R/L mode
Manual
The other eye button is used to manually switch between right
and left eyes in continuous measurement mode.
Full Auto
Right and left eye switching is automatically performed in continu-
ous measurement mode before measurement.
Auto (RL)
Right and left eye switching is automatically performed in continu
-
ous measurement mode, but no measurement is performed.
Touch Measure
OFF Touch measurement is not performed in Manual mode.
ON Touch measurement is performed in Manual mode.
IOP Adjustment
OFF Disables the intraocular pressure correction function.
ON Enables the intraocular pressure correction function.
Center CCT Base
0-999
Set by number display.
Sets the central cornea thickness base value.
(Used when the corrected intraocular pressure value is “ON”.)
545
520 545
Adjustment Coefficient
0-9999
Set by number display
Sets the adjustment coefficient.
500 means 0.050 (mmHg/μm) in adjustment coefficient.
(Used when the corrected intraocular pressure value is “ON”.)
500
120 500
IOL LED Brightness
0-100
Set by number display
Sets brightness of alignment dot in IOL mode. 28
Meas. Data recording way
Data on credibility
Measurement data is displayed in the order
from low to high credibility.
Data without error The measurement data without error is displayed.
Data with error All the measurement data (including data with error) is displayed.
Descriptions Options Details
Initial value
(Region)
A C E D J G
Auto Contact Avoidance
OFF
The measurement nozzle does not stop automatically when it
approaches the patient.
ON
Automatically stops operation when the nozzle approaches the
patient.
Meas. Count change mode
1×
Sets the Meas. Count change mode at the start of measurement
to one-time measurement.
Multi
Sets the Meas. Count change mode at the start of measurement
to the preset number of measurements.
Descriptions Options Details
Initial value
(Region)
A C E D J G
Cont. cycle (TONO)
1-10
Set by number display
The number of continuous measurements (TONO). 3
Show Tono average
OFF Tono average value is not displayed.
ON Tono average value is displayed.
Show Adj value
OFF Adjusted measurement value is not displayed.
ON Adjusted measurement value is displayed.
Tono display Unit
mmHg Display in mmHg
digit Display in digit
hPa Display in hPa
Torr Display in Torr
Press average Mode
Integer Displays the average intraocular pressure value as an integer.
Real Displays the average intraocular pressure value as a real number.
Low Credibility Measure
No Including The value of low credibility is excluded from count cycle.
Including The value of low credibility is included to count cycle.