139
GENERAL INFORMATION ON USAGE AND MAINTENANCE
Guidance and manufacturer's declaration - electromagnetic immunity
The TRK-3 OMNIA is intended for use in the electromagnetic environment specified below.
The customer or the user of the TRK-3 OMNIA should assure that it is used in such an environment.
Immunity test Test level Compliance level Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
6 Vrms
Within ISM band
and amateur radio
band of
150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
Proximity
electromagnetic
field from radio
communication
equipment
a)
3 Vrms
150 kHz to 80 MHz
6 Vrms
Within ISM band
and amateur radio
band of
150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
Proximity
electromagnetic
field from radio
communication
equipment
a)
Portable and mobile RF communications
equipment should be used no closer to any
part of the TRK-3 OMNIA, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d =
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer, d is the recom
-
mended separation distance in meters (m),
and E is the radiation electromagnetic field
level in volt/meter (V/m).
Proximity magnetic
fields
IEC 61000-4-39
30 kHz CW
8 A/m
134.2 kHz
PM 2.1 kHz 50%
65 A/m
13.56 MHz
PM 50kHz 50%
7.5 A/m
30 kHz CW
8 A/m
134.2 kHz
PM 2.1 kHz 50%
65 A/m
13.56 MHz
PM 50kHz 50%
7.5 A/m
The exterior surface of the TRK-3 OMNIA
should be kept at least 0.15 m from RF emit-
ters such as RFID readers.
NOTE 1
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a) The table below shows the proximity electromagnetic field from radio communication equipment.
P
6
E
Test frequency
[MHz]
Band [MHz] Equipment Modulation
Maximum output
[W]
Distance [m]
Immunity test
value [V/m]
385 380–390 TETRA 400
Pulse modulation
18 Hz
1.8 0.3 27
450 430–470 GMRS 460 FRS 460
FM ±5 kHz 1kHz
sine
2 0.3 28
710
704–787 LTE Band 13, 17
Pulse modulation
217 Hz
0.2 0.3 9745
780
810
800–960
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5
Pulse modulation
18 Hz
2 0.3 28
870
930
1720
1700–1990
GSM 1800
CDMA1900
GSM 1900
DECT
LTE Band 1, 3, 4, 25
UMTS
Pulse modulation
217Hz
2 0.3 28
1845
1970
2450 2400–2570
Bluetooth
WLAN 802.11 b/g/n
RFID 2450
LTE Band7
Pulse modulation
217Hz
2 0.3 28
5240
5100–5800 WLAN 802.11 a/n
Pulse modulation
217Hz
0.2 0.3 95500
5785