RER.
Engineer
Troubleshooting
Low
Suction
Symptom:
No
suction
present
at
handpiece
tip
Is
handpiece
tip
biocked?
Remove
obstruction
from
tip
No
Is
handpiece
manifold
tubing
kinked
and/or
is
it
properly
connected?
Yes
Remove
kinks
from
manifold
tubing
and
re-connect
tubing.
ts
suction
canister
installed
correctly?
if
liner
is
used,
is
suction
canister
finer
installed
correctly?
Install
suction
canister
correctly.
ts
suction canister
full?
install
suction
canister
liner
correctly.
Install
new
suction
canister
Is
contamination
guard
tube
kinked?
Remove
kinks
from
contamination
guard
tube.
No
Continued
on
next
page.
10-20
CUSA
EXcel-8
System
Service
Manual