Electrical
Requirements
Fluidic
System
Cooling
System
Cooling
water
flow
is
35-50
cc/min
Irrigation
Rate
The
irrigation
rate
display
shows
digits
1-10
The
irrigation
rate
is
approximately
1-10
cc/min,
normal;
25-30
cc/min,
Fast
Flush
Suction
System
Up
to
660
mm
mercury
at
the
pump
intake
at
sea
level
The
suction
level
will
be
lower
at
higher
altitudes.
Electrical
Requirements
Input
Power
Source
Power
Rating
The
CUSA
Excel
220-240
Volt
(default)
system
operates
on
one
of
the
following
95-120
Volt
voltage
ranges
Frequencies
50
Hz
+
2
Hz
60
Hz
+
2
Hz
Line
fusing
resides
in
fuse
receptacles
on
the
rear
panel.
The
fuse
values
for
220-240
Volt
operation
are
T5A;
for
95-120
Volt
operation,
T10A.
Refer
to
Section
11,
Replacement
Procedures
for
instructions
on
replacing
the
fuses.
Power
Cords
Cord
Harmonized
type
three
conductor
cord
Connector
Continental
European
connector
If
a
facility
intends
to
use
the
CUSA
EXcel
system
in
an
operating
room
with
another
type
of
receptacle,
the
facility
is
responsible
for
replacing
the
connector
on
the
power
cord.
The
replacement
connector
must
be
properly
grounded.
CUSA
EXcel-8
System
Service
Manual
93
=
|
gg
$и0129119946
[2911991

Table of Contents