How
the
TISSUE
Select
Feature
Works
TISSUE
Select
Operation
Power
is
interrupted.
The
ultrasonic
generator
provides
less
reserve
power
to
the
handpiece:
*
The
tip
still
fragments
“softer”
tissue
easily.
*
But,
when
it
encounters
“stronger”
tissue,
the
power
the
tip
receives
is
no
longer
enough
to
fragment
the
tissue.
The
tip
stalls.
At
increased
selectivity
settings
(+
to
++++),
suction
and
the
tip
cross-
sectional
area
remain
the
same,
and
stroke
amplitude
and
tip
acceleration
decrease
slightly:
Table
6-2.
Fragmentation
performance
vs.
TISSUE
Select
settings
Setting
Fragmentation
Rate
Std
and
+
maximum
++
and
+++
slightly
decreased
rate,
increased
tissue
selectivity
and
tactile
feedback
++++
slowest
rate,
maximum
selectivity
and
tactile
feedback
Benefits
of
TISSUE
Select
The
TISSUE
Select
feature
presents
several
benefits:
*
Maintains
fragmentation
ability.
*
Reduces
(automatically)
fragmentation
rate.
ο
Provides
maximum
tissue
selectivity.
*
Gives
surgeon
superior
tactile
feedback.
*
Gives
surgeon
greater control
and
precision
when
dissecting
near
delicate
structures.