Environment
Explosion
Hazard
Do
not
use
the
CUSA
EXcel
system
in
the
presence
of
flammable
anesthetics
or
any
potentially
explosive
or
flammable
atmosphere.
Low
Frequency
Leakage
Enclosure
leakage
current
Patient
leakage
current
Duty
Cycle
<300
pA
<100
pA
Under
maximum
loading
conditions,
the
CUSA
EXcel console
is
suitable
for
ultrasonics
activation
times
of
10
minutes
on,
5
minutes
off.
Environment
Operating
temperature
10°
C
to
40°
C
range
Operating
humidity
range
15
to
80%,
relative
humidity,
non-condensing
Operating
atmospheric
70
kPa
to
106
kPa
pressure
range
Storage
and
shipping
-34°
C
to
65°
C
temperature
Storage
and
shipping
25
to
85%
relative
humidity,
non-condensing
humidity
The
system
requires
a
minimum
of
one
hour
exposure
at
its
operating
temperature
range
before
you
use
it.
!
|
5-4