CHAPTER 3AUTOMATED EXTERNAL DEFIBRILLATOR (AED) OPERATION
3–8 www.zoll.com 9650-0904-01 Rev. N
Repeat Analysis
After performing CPR for the configured period of CPR, the unit will restart ECG analysis
automatically.
Note: Reanalysis of the ECG rhythm is inhibited for 3 seconds after a shock.
Continue Patient Care
Continue patient care according to medical protocols.
Operating Messages
The unit uses both audio and visual prompts to present critical information to operators. The
following information describes the unit default configuration. If your device has been custom
configured, some of the information may be different.
There are 10 voice prompts used in AED mode. Most of these prompts are accompanied by a
message displayed on the monitor. The voice prompts are given only once, but the monitor
continues to display the message until new action is taken by the operator or the device status
changes.
The unit will alternately display two different messages in the same field of the display when
two conditions are detected at the same time. For example, a LOW BATTERY message may
alternately display on the same line of the monitor as the CHECK PADS message.
Audio and Display Messages
The display messages and voice prompts that can occur during AED operation are described
below.
ATTACH PADS
If the unit is powered on without therapy pads connected to the patient, the ATTACH PADS
message is announced and displayed.
ANALYZING ECG/STAND CLEAR
The ANALYZING ECG message is displayed and the STAND CLEAR message is displayed and
announced when the ECG analysis starts automatically or after pressing the ANALYZE
softkey. They indicate that an active ECG analysis is in progress.
CHARGING XXXJ
ECG analysis is still in progress and a potentially shockable rhythm has been detected. The
current charge level is displayed.
SHOCK ADVISED
A shockable rhythm has been detected and defibrillation is advised. The selected energy level
has not yet been reached. The current energy level is displayed.

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB