CHAPTER 4MANUAL DEFIBRILLATION
4–2 www.zoll.com 9650-0904-01 Rev. N
Begin CPR Following Local Medical Protocols
Request additional assistance.
1 Select ON
The unit beeps 4 times to indicate that it has passed the power on self test, and the green AED
label and the ON mode selector light up on the front panel.
Press the Manual Mode softkey on the front panel of the unit to enter the Manual mode of
operation.
The message CONFIRM MANUAL MODE and the Confirm softkey will be displayed. Press
the Confirm softkey to enter the Manual mode of operation. If the Confirm softkey is not
pressed within 10 seconds, the unit will revert back to AED operation.
Note: If the unit has been configured not to confirm the transition from AED mode to
Manual mode, the unit will immediately enter Manual mode.
The control buttons for Manual mode light up on the front panel and the unit automatically
defaults to 120 joules or the preconfigured first shock energy selection.
Note: Defibrillator PADDLES are selected as the ECG source when the instrument is
changed to Manual mode with paddles connected to the OneStep cable.
Energy Select
Look at the Display and verify the energy is appropriate. Unless internal handles are connected
to the OneStep cable, the default energy selections for adult patients are:
Shock 1 - 120 joules
Shock 2 - 150 joules
Shock 3 - 200 joules
OFF
ON
AED

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB