Transferring Device Check and Activity Log Files Through Wi-Fi (Optional)
9650-0912-01 Rev. N ZOLL R Series Operator’s Guide 11–7
Related Wi-Fi Messages
Informational Message Description
INITIALIZING NETWORK A transfer has been initiated.
CONNECTING TO THE NETWORK The association with the host network is being
established.
CONNECTING TO SERVER The unit is connecting to a remote system.
TRANSFER IN PROGRESS The data transfer is in progress.
WAITING FOR SERVER RESPONSE The unit is waiting for the final acknowledgment from
a remote system.
TRANSFER COMPLETE
XX Was Sent
XX=file type (Full Disclosure, Activity
Log, or Device check)
The data transfer is complete (and lists the file type
that was sent).
Error Message Description/Action
TRANSFER FAILED
Wi-Fi Card Not Detected
Verify Installation of Wi-Fi Card
The transfer failed because no Wi-Fi card is installed
in the CF slot, or the card was ejected from the slot
during transmission.
Action: Make sure that the correct Wi-Fi card is
properly installed in the unit.
TRANSFER FAILED
Wi-fi Card Not Configured
Contact Network Administrator
The Wi-Fi card’s configuration data is corrupt or blank.
Action: Verify the configuration settings on the Wi-Fi
card. Contact the Network Administrator for
assistance.
TRANSFER FAILED
Invalid Wi-Fi configuration
Error Number: N
Contact Network Administrator
The transfer failed because the Wi-Fi configuration
was illegal. N indicates one of following error codes:
N = 1: Local Static IP settings
Action:
1) If running in Static IP mode, make sure that the
relevant static IP addresses are not 0.0.0.0.
2) Make sure the subnet mask is in the correct format.
N = 2: Server Static IP address
Action: With DNS set to No, make sure that the IP
addresses for the Full Disclosure Server and Defib
History Server are not 0.0.0.0.
N = 3: Server Name
Action: With DNS set to Yes, make sure that valid
names exist for the Full Disclosure Server and the
Defib History Server.
N = 4: DNS IP address
Action: With DHCP set to No and DNS set to Yes,
make sure that the DNS IP Address is not 0.0.0.0.
N = 5: SSID
Action: Make sure that there is at least one valid
SSID for each desired mode (infrastructure and data
management server).

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB