Manual Defibrillator Testing
9650-0904-01 Rev. N ZOLL R Series Operator’s Guide 13–7
10 mm ±1.0 mm high
Code Readiness Log
Each automatic and manual defibrillation test result is stored in internal non-volatile memory,
the Code Log. A total of 1000 Code Readiness test records can be stored in internal memory.
When the Code Readiness Log is full, the oldest records are erased on a first-in-first-out (FIFO)
basis.
The Code Readiness Log can be transferred to an external computing device (see “Transferring
Files to an External Device” on page 12-1).
If configured, the R Series prints a Code Readiness Test Report following the completion of
each automatic defibrillator test.
To print the Code Readiness Log:
1. Turn the Mode Selector to ON.
2. Press the Manual Mode softkey.
Note: If the CONFIRM MANUAL MODE message displays, press the Confirm softkey to
enter the Manual mode of operation.
3. Press the Report Data softkey.
4. Press the Test Log softkey.
A menu appears with the print options.

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB

Related product manuals