CHAPTER 14 TROUBLESHOOTING
14–4 www.zoll.com 9650-0904-01 Rev. N
Recorder
Pacer
Symptom Recommended Action
1. CHECK RECORDER message
Recorder out of paper.
Remove paper, check paper type, check recorder for
paper jam, and reload paper.
Recorder door is open.
Paper is loaded upside down.
2. Recorder makes stuttering sound
when activated.
Check recorder for paper jam.
3. Light or poor quality printing
Ensure correct paper is in use.
Ensure paper is installed grid side against recorder print
head.
Recorder print head requires cleaning.
4. Summary Report will not print
when Report/Print Chart softkeys
are activated.
15 seconds have not elapsed since one of the events
that trigger Summary Report to record have occurred.
Wait 15 seconds, then try again.
Symptom Recommended Action
1. CHECK PADS message
Ensure therapy electrodes are connected to the
OneStep cable.
Ensure electrode gel is not dry. Replace therapy
electrodes if necessary.
Ensure good electrode-to-patient contact.
Check integrity of OneStep cable by plugging into test
connector. CHECK PADS should disappear.
2. No stimulus marker ( ) is
present on the ECG trace.
Ensure unit is in PACER mode.
Ensure PACER RATE (ppm) is set greater than patient
heart rate.
3. No ventricular capture beat
appears after stimulus marker on
ECG display.
Check patient’s pulse.
Increase output current.
Ensure therapy electrodes are making good contact
with the patient.
Select different ECG Lead configuration.
Review therapy electrode placement.

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB

Related product manuals