9650-0912-01 Rev. N ZOLL R Series Operator’s Guide 12–1
Chapter 12
Maintenance
Defibrillator equipment must be maintained to be ready for immediate use. The defibrillator
should be tested daily. The R Series defibrillator gives you two readiness testing options:
Automatic
Manual
Note: Both automatic and manual test results are automatically recorded to internal memory.
When the R Series device ships from ZOLL, the Code Readiness indicator may show a red
“X.” A manual readiness test should be performed (in addition to other site-specific tests such
as HiPot and Leakage) prior to putting the device into service. Follow the procedure on page
12-3, and verify that the Code Readiness indicator displays a green “.”
Maintenance Frequency
Visual Inspection Once per day, inspect Code Readiness indicator.
Code Readiness Test
(Automatic)
Once per day, unless configured Off, in which case, perform manual
defibrillator testing daily.
Manual Defibrillator
Testing
Once per week; daily if Code Readiness Test is configured Off.

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB