CHAPTER 12 MAINTENANCE
12–6 www.zoll.com 9650-0912-01 Rev. N
The messages CHECK PADS and POOR PAD CONTACT appear alternately. The pace alarm
sounds, and the Clear Pace Alarm softkey flashes.
6. Reconnect the OneStep cable, and press the Clear Pace Alarm softkey.
The messages CHECK PADS and POOR PAD CONTACT disappear, and the alarm tone
stops.
Recorder Check
1. Check the printer for an adequate supply of paper, then press the RECORDER button.
2. Press and hold the SIZE button for at least 2 seconds.
A calibration pulse of 1 mV appears on the display while the button is held. The amplitude
of the calibration pulse is independent of the SIZE setting.
3. Inspect the recorder waveform for uniformity and darkness.
4. Inspect for uniformity of annotated characters and completeness of words.
5. Check the printer speed by verifying that the resulting waveform is:
2.5 mm ±0.5 mm wide
10 mm ±1.0 mm high
Code Readiness Log
Each automatic and manual defibrillation test result is stored in internal non-volatile memory,
the Code Log. A total of 1000 Code Readiness test records can be stored in internal memory.
When the Code Readiness Log is full, the oldest records are erased on a first-in-first-out (FIFO)
basis.
The Code Readiness Log can be transferred to an external computing device (see “Transferring
Files to an External Device” on page 11-1).
If configured, the R Series prints a Code Readiness Test Report following the completion of
each automatic defibrillator test.
To print the Code Readiness Log:
1. In MONITOR mode, press the Report Data softkey, then the Test Log softkey.
A menu appears with the print options.
2. To print a specific test, use the Prev Test and Next Test softkeys to select the test, then press
Print Test, or press Print Test Log to print the log of all of the tests.

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory