CHAPTER 6REAL CPR HELP
6–4 www.zoll.com 9650-0912-01 Rev. N
By default, the FULLY RELEASE text prompt is not enabled.
This feature is unavailable while using Pediatric CPR electrodes.
CPR Voice Prompts (Adult only)
The R Series unit can be configured to issue voice prompts related to the depth of chest
compressions as feedback to rescuers performing CPR. Two voice prompts are available for
this purpose:
Push Harder
Good Compressions
When chest compressions are detected but their depth is consistently less than 1.5 or 2 inches
(3.8 or 5 cm), depending on the configuration, the defibrillator will issue the prompt “Push
Harder” every 15 seconds. If the rescuer responds by increasing compression depth to more
than 1.5 or 2 inches (3.8 or 5 cm), depending on the configuration, on a consistent basis, the
unit will issue a “Good Compressions” prompt.
See the R Series Configuration Guide for information on enabling/disabling CPR voice
prompts.
CPR Voice prompts are unavailable while using Pediatric CPR electrodes.
Chest Compressions Bar Graph
The R Series unit can display a CPR compression bar graph computed from the CPR sensor
signals. This bar graph, representing depth of compression, is presented on a displacement scale
with a reference marker at 1.5 or 2.0 inches, depending on the configuration. When the full
width of the trace is visible, the unit displays a minimum of 12 seconds of compression data.
Displaying the CPR Bar Graph
To display the CPR displacement bar graph in the Trace 2 or 3 position:
1. Press the Options softkey, then press Traces.
2. Press either the Trace 2 or Trace 3 softkey.
3. Press CPR.
Note: The CPR softkey appears only when OneStep CPR or OneStep Complete electrodes
are in use.

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB