APPENDIX ASPECIFICATIONS
A–2 www.zoll.com 9650-0904-01 Rev. N
Defibrillator Specifications
General
Size
(height • width •
length)
8.2 in. • 10.5 in. • 12.5 in. with handle or 10.0 in. without handle
20.8 cm • 26.7 cm • 31.7 cm with handle or 25.4 cm without handle
Weight 13.6 lb (6.17 kg) with OneStep cable and battery pack
15.2 lb (6.89 kg) with paddles
Power (for
R Series ALS, BLS
and Plus models)
Battery: Rechargeable lithium ion battery pack
AC Power Requirements: 100-120 Vrms, 50/60 Hz
220-240 Vrms, 50 Hz
AC Power Consumption: 275 VA maximum
Device
classification
Class I and internally powered per EN 60601-1.
Design standards Meets or exceeds applicable requirements of UL 60601, AAMI DF80, IEC
60601-2-4, EN 60601-2-25, and EN 60601-2-27.
Patient safety All patient connections are electrically isolated.
Environmental
Operating
temperature
0°C to 40°C (32°F to 104°F)
Storage and
shipping
temperature
–20°C to 60°C (-4°F to 140°F)
Humidity 5% to 95% relative humidity, noncondensing
Vibration IEC 68-2-6 and IEC 68-2-34
Shock IEC 68-2-27, 50 g 6ms half sine
Operating
pressure
594 to 1060 millibars (-1253 to 14046 ft.)
Particle and water
ingress
IEC 529, IP 22
Electromagnetic
compatibility
(EMC)
CISPR 11 Class B - radiated and conducted emissions
Electromagnetic
immunity
AAMI DF80; EN 61000-4-3 to 10 V/m
Electrostatic
discharge
AAMI DF80; EN 61000-4-2
Conducted
susceptibility