R Series Service Manual
1-20 9650-0903-01 Rev. L
10.0 Heart Rate Alarm Test
Tools Needed:Impulse 4000.
Do this... Observe this... Pass Fail/NA
10.1 Turn the selector switch to MONITOR mode( for
AED units turn to ON.)
Connect the ECG leads to the Impulse 4000.
Set the simulator to 120 BPM and the
defibrillator to lead II.
Lead II waveform displayed on monitor.
NSR ECG at 120 BPM +/- 2 displayed.
ooo
10.2 Press ALARMS softkey. The alarm menu displays.
ooo
10.3 Press NEXT PARAM softkey until ECG HR
displays.
Cursor scrolls through parameters.
ooo
10.4 Press CHANGE VALUE softkey.
ooo
10.5 Press INC> softkey for state. Cursor scrolls through ENABLE, AUTO and DISABLE.
ooo
10.6 Press DEC< softkey for state. Cursor scrolls through ENABLE, DISABLE, AND AUTO.
ooo
10.7 Press INC> softkey until ENABLE displays. ENABLE displays.
ooo
10.8 Press ENTER softkey.
ooo
10.9 Press NEXT FIELD softkey to select the heart
rate limit.
Cursor scrolls to Low field.
ooo
10.10 Press CHANGE VALUE softkey and press
INC> or DEC< to set low heart rate limit to 30.
ooo
10.11 Press ENTER softkey.
ooo
10.12 Press NEXT FIELD softkey to select the high
heart rate limit.
Cursor scrolls to High field.
ooo
10.13 Press CHANGE VALUE and press INC< or
DEC> softkey and set high heart rate limit to
150.
ooo
10.14 Press ENTER softkey, then press RETURN
softkey.
ooo
10.15 Press ALARM SUSPEND button. No alarm sounds.
ooo

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB

Related product manuals