Page #222 background imageLoading...
Page #222 background image
CHAPTER 3AUTOMATED EXTERNAL DEFIBRILLATOR (AED) OPERATION
3–4 www.zoll.com 9650-0904-01 Rev. N
2. Roll the pad smoothly from the applied edge to the other, being careful not to trap any air
pockets between the gel and skin.
Note: If it is not possible to place the “BACK” pad on the patient’s back, the pads should be
placed in the standard apex-sternum positions. Effective defibrillation will result, but
pacing will usually be less effective.
1 Select ON
The unit beeps 4 times to indicate that it has passed the power on self test, and the green AED
label and the ON mode selector light up on the front panel.
If no hands-free therapy electrodes have been attached to the patient and connected to the
R Series, the ATTACH PADS message and voice prompt will be issued.
Energy Select
The default energy selections for adult patients are:
Shock 1 - 120 joules
Pad
Skin
OFF
ON
AED
ATTACH PADS
Manual
Mode
ECG
120J SEL
Analyze
ATTACH
PADS

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB