EasyManuals Logo

ZOLL R Series Service Manual

ZOLL R Series
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #339 background imageLoading...
Page #339 background image
Defibrillator Specifications
9650-0904-01 Rev. N ZOLL R Series Operator’s Guide A–7
Battery Pack Specifications
Type Rechargeable lithium ion
Weight 1.7 lb (0.77 kg)
Nominal voltage 10.6 V
Recharge time 4 hours or less within R Series.
Operating time For a new, fully charged battery at 20°C:
100 defibrillator discharges at maximum energy (200 joules), or
4 hours of continuous ECG monitoring, or
3.5 hours of continuous ECG monitoring and pacing at 60 mA, 80
pulses per minute
Low battery indicator The message LOW BATTERY is displayed on the screen when there
is approximately 15 minutes of ECG monitoring time left on the
battery. Two-beep low battery tone sounds once a minute until just
before shutdown when the unit beeps twice every 2 seconds.
The time from display of the message LOW BATTERY or REPLACE
BATTERY until the defibrillator shuts down varies depending on the
battery age and condition.
Battery Shelf Life 3 months before retest and recharge

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB