R Series Service Manual
9650-0903-01 Rev. L 1-23
11.3 Select energy level of 100J and press the
CHARGE button.
The charge time is >2 second and <10 seconds and SELECT 30J FOR TEST
is displayed.
oo
11.4 Press the SHOCK button. Unit does not discharge. DEFIB PAD SHORT message displays.
oo
11.5 Set energy level to 30J. Unit internally discharges.
oo
11.6 Press the CHARGE button. Unit charges to 30J and displays DEFIB 30J READY. The charge ready tone
sounds.
oo
11.7 Press and hold SHOCK button. Unit discharges. 30J TEST OK message displays. The message at the top of
the printed strip chart reads as follows:
30 JOULES TEST OK. TEST_CUR=10-14A DEFIB_IMPED=0.
The impedance value may range from 0 to 5.
oo
Record your results on the Maintenance Test Checklist.
Do this... Observe this... Pass/Fail

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB

Related product manuals