R Series Service Manual
9650-0903-01 Rev. L 1-37
Note: The calibration gas mixture and regulator are available from most medical gas supply companies. You can also try Scott Medical products at
www.scottmedicalproducts.com.
21.10 Set the Gas Balance settings of the
CAPNOSTAT 5 CO
2
Sensor to that of the
calibration gas mixture (N
2
, N
2
O, or He).
The default gas balance is N
2
.
21.11 Allow a few seconds for the gas mixture to
stabilize and observe the CO2 Percent value.
The expected value is 5% ± 0.2%.
ooo
21.12 Press the Return softkey to return to the main
EtCO
2
Calibration screen.
21.13 Turn the device off when calibration is
complete.
Record your results on the Maintenance Test Checklist.
Do this... Observe this... Pass/Fail/NA

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB

Related product manuals