EasyManuals Logo

ZOLL R Series Service Manual

ZOLL R Series
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
R Series Service Manual
9650-0903-01 Rev. L 2-7
ENERGY INCREMENTED Defib energy has been automatically incremented to the
next configured level after shock 1 or 2 has been delivered
and the unit is configured for Basic Energy Auto
Escalation.
ERASING REPORT The unit is erasing the selected report data.
FULL DISCLS STOPPED Full disclosure data exceeds the storage capacity: 4 hours
or 32Mb.
IF NO PULSE Message displayed upon entered CPR period specified by
AHA protocol.
INSERT CARD No card detected during manual or semi-automatic modes.
Re-insert CF Card or replace communication
module.
INT. DUMP OVERLOAD More than 15 internal discharges in 5 minutes. Unit needs to cool down, wait approximately
5 minutes.
INVALID GAS TEMP:
RETRY
The operator has attempted to set a gas temperature
outside the Capnostat’s operating range.
Recalibrate EtCO
2
, verify range.
LOW BATTERY Low battery.
Replace battery pack with a fully charged
battery pack.
Plug unit into AC mains.
NIBP ARTIFACT The unit is unable to detect systolic, diastolic or mean
blood pressure due to excessive motion or vibration.
Take a single blood pressure measurement.
Keep patient as still as possible.
Insulate patient, cuff and hose from vibrations
as much as possible.
NIBP COMM ERR 259 Framing, parity, or fifo error. Replace parameter power supply or NIBP module.
NIBP COMM ERR 260 Received invalid packets. Replace parameter power supply or NIBP module.
NIBP COMM ERR 261 Device not taking action request. Replace parameter power supply or NIBP module.
NIBP COMM ERR 262 No reply from device. Replace parameter power supply or NIBP module.
Error Message Explanation User
Advisory
Technical Action

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB

Related product manuals