9. Lift up on the recorder tray from the rear of the
R Series to release it.
10.Rotate device so the display is facing you. Lift up on the
paper tray to expose cables. Disconnect the recorder
interconnect cable from the print head and the motor/
sensor board.
Note: During reassembly, observe the labeling on the cable to ensure you
are connecting the cable correctly.
Recorder
cables
interconnect