4.2.2 Operations in Pulse Source Mode
4-13
a. When the trigger mode is AUTO:
The measurement and the pulse cycle repeat at intervals of the specified period time.
If the measurement does not finish in the specified period, the pulse width does not change but
the pulse cycle is extended.
If the source value or the base value is changed during pulse generation, the pulse source
operation stops and restarts with the new source value or base value.
b. When the trigger mode is HOLD:
The pulse generation starts when the source delay time has passed after trigger input.
The measurement starts when the measurement delay time has passed after trigger input.
Trigger input during the period time is ignored.
c. When setting to Operate or changing the range
Pulse generation starts when the Operate processing time has passed after setting to Operate.
If the source change entails the range change, the base value and pulse value are set in the same
range.
d. In the Standby or Suspended status:
The measurement is not executed in the Standby or Suspend status.