6253 DC Voltage Current Source/Monitor Operation Manual
4.2.3 Operations in Sweep Source Mode
4-15
1. Setting the sweep type
• Select DC Sweep or Pulse Sweep by using the soft key MODE.
• Select Linear, MLinear, Log, Random or Fixed by using the soft key SWPType.
Pulse
sweep
PLSSWP
Linear sweep
Linear
Generates staircase pulse waveforms of the step value
between the start value and the stop value.
Kth output value = Start value +
(K - 1)
× (Step value)
Number of steps = (|Stop value – Start value|) /
(Step value) +1
Multi-Slope Linear
Sweep
MLinear
Generates staircase pulse waveforms of different slopes
by specifying the number of slopes between 2 and 4.
1st slope:
Kth output value = First value +
(K - 1)
× (1st step value)
Number of steps = (|Second value - First value|) /
(1st step value) +1
2nd slope:
Kth output value = Second value +
(K - 1)
× (2st step value)
Number of steps = (|Last value - Second value|) /
(2nd step value) +1
Log sweep
Log
Generates staircase pulse waveforms logarithmically
divided by the number of steps by decade.
Kth output value = Start value
× S
S = 10
(K - 1) / n
n: Number of partitions by decade
(1, 2, 5, 10, 25, 50)
• Start value, Stop value 0
• |Start value| | Stop value|
• Start value polarity = Stop value polarity
Random sweep
Random
Generates pulse waveforms of the source values stored
in the memory from the start address to the stop
address.
Fixed level sweep
Fixed
Generates pulse waveforms of the specified level by the
specified sampling count.
Table 4-3 Operations in Sweep Source Mode (2 of 2)
Sweep type Operational Descriptions Waveform
Step value
Stop value
Start value
Bias value
Base value
Last value
First value
Bias value
2nd step value
1st slope
2nd slope
1st step value
Second
value
Specify the number of slopes
(2 to 4).
Base value
Stop value
Start value
Bias value
Base value
Bias value
Start address
Stop address
Base value
Bias value
Level value
Sampling count
Base value