6253 DC Voltage Current Source/Monitor Operation Manual
4.2.3 Operations in Sweep Source Mode
4-16
2. Changing the sweep measurement parameters
The sweep measurement parameters are basically changeable only in the Standby or Suspend status,
however the following parameters are changeable during sweep stop in the Operate status.
• Time parameter
• Hold time
• Source delay time
• Measurement delay time
• Pulse width
• Period time
• Start address and stop address for random sweep
(These are changeable in a range between the start address and the stop address range when
switching from the Standby status to Operate or Suspend.)
• Sweep function parameters
• Repeat count
• Reverse function ON or OFF
• Return to Bias ON or OFF
• Measurement auto range ON or OFF
• Measurement ON or OFF
• Measurement integration time
• Selecting COMPLETE OUT or SYNC OUT external control signal output function