6253 DC Voltage Current Source/Monitor Operation Manual
4.2.8 Source and Measurement Timing
4-56
4.2.8.6 Variable Slew Rate
The variable slew rate is used to set the rise time and fall time for voltage source to any values.
• There are the following four ranges for the variable slew rate and the output response changes
automatically.
NOTE:
1. During Operate, the setting value can be changed only in the same range.
When range change occurs, set the output status to Standby.
2. The Auto Range for source and measurement is not available.
3. When the Slew Rate Indiv is ON, the range corresponding to the slower setting value is
applied, and the fastest value in that range is set as the faster value.
(Example) Slew Rise: 100.0 V/s and Slew Fall: 5.000 kV/s
Slew rate range: 2
Slew Rise: 100 .0 V/s
Slew Fall: 999.9 V/s
Range Setting value Response Setting accuracy
1 10.00 to 99.99 V/s Slow (20 % + 500 digits)
2 100.0 to 999.9 V/s Slow Pure resistive load
3 1.000 to 9.999 kV/s Fast IL: 3 mA range or higher
4 10.00 to 99.99 kV/s Fast When the limit does not activate
7) Slew Rate SW
Variable slew rate ON or OFF
8) Slew Rise
Rise time setting
9) Slew Fall
Fall time setting
10) Slew Rate Indiv
Rise and fall time individual setting
ON or OFF