6253 DC Voltage Current Source/Monitor Operation Manual
4.2.11 Multiple Unit Operation
4-68
From the above, considering the Tsync time, make the relevant settings as follows:
1. Tds1 Tds2 + Tsync (5 s)
2. Td1 Td2 + Tsync (5 s)
3. Tp1 Tp2 + TA (1 ms)
(Tp1 1 ms, Tp2 500 s)
4. Th1 Th2 - Th(ext) - Tp(ext)
(Th (ext) min = 0.1 ms, Tp (ext) min = 1 ms)
Example) Tds1 = 35 s
Td1 = 55 s
Tp1 = 1.5 ms
Th1 = 1.1 ms
Tds2 = 30 s
Td2 = 50 s
Tp2 = 500 s
Th2 = 0 ms