6253 DC Voltage Current Source/Monitor Operation Manual
5.8.3 Remote Command List
5-37
Sweep Random
sweep
memory data
N [adr]
P
Random sweep memory data setting starts from the
N command and completes at the P command.
N<adr>, SVR<n>, SOV<data1>, SOV
<data2>, ... ,P (for voltage setting)
N<adr>, SIR<n>, SOI<data1>, SOI
<data2>, ... ,P (for current setting)
Yes No
adr: Memory address (0 to 19999)
data1: Voltage or current source value at adr
data2:
Voltage or current source value
at
adr+1
0
0
*6
N? [adr] Response: N<adr>, SVR<n>, SOV<data>, P
(For voltage source value)
N<adr>, SIR<n>, SOI<data>, P
(For current source value)
adr: <ddddd>
n: <d>
data: <d.ddddEd>
*1
Yes Yes
NP? Query of the random sweep memory setting status
Response: 0 ...Random sweep memory setting complete
1
... Random sweep memory is being set.
0YesYes
RSAV Saves the random sweep data.
Yes No
RLOD Loads the random sweep data.
Yes No
RCLR Initializes the random sweep data.
(Data saved in the memory is not initialized.)
Yes No
Pulse sweep
base value
BS [data] data: Pulse sweep base value
0Yes
BS? Response: BS<d.dddddEd>
*1
Yes Yes
Bias value SB [data] data: Bias value
0Yes
SB? Response: SB<d.dddddEd>
*1
Yes Yes
*1: The decimal point position in response varies depending on the setting value.
For the source value, limit value, and time parameter setting ranges, refer to the specifications.
*6: Not initialized by the RINI or *RST command.
Item Command Description
Default Operation possibility
Power
ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS
NOTE:
1. When no source range is specified, the
optimal range is set.
2. Source values different from the present
source function cannot be set.