6253 DC Voltage Current Source/Monitor Operation Manual
7.5.6 Current Measurement Calibration (3mA to 300 mA)
7-16
9. Finish the DMM data input mode by XNXT.
10. Move to the zero calibration data fine adjustment mode by XADJ.
11. Check the zero calibration value.
Fine-adjust the calibration coefficient by XUP and XDN.
12. Move to the full-scale calibration data fine adjustment mode by XNXT.
13. Check the full-scale calibration value.
Fine-adjust the calibration coefficient by XUP and XDN.
14. Move to the next step
When changing the current calibration mode: XNXT
When moving to current measurement calibration: XNXT XIM
7.5.6 Current Measurement Calibration (3A to300 mA)
1. Move to the current measurement calibration mode by XIM.
2. Select the range.
3 A : XR-2
30 A : XR-1
300 A : XR0
3 mA : XR1
30 mA : XR2
300 mA : XR3
3. From this step, perform the same procedure as described in Section 7.5.3,
“Voltage Measurement Calibration.”