6253 DC Voltage Current Source/Monitor Operation Manual
7.5.7 Current Measurement Calibration (2 A)
7-17
7.5.7 Current Measurement Calibration (2 A)
1. Make a connection in reference to Figure 7-1 (c) and set the DMM measurement
function to DCV.
2. Move to the voltage calibration mode by XIM.
Select the range.
2 A : XR4
3. Enter the DMM data input mode by XDAT.
4. Obtain the DMM reading value by using the following formula:
data = DMM reading value (V) / Standard resistance value ()
5. Set the calculated data by XD data.
6. Read and check the measurement data.
7. Move to the full-scale calibration mode by XNXT.
8. Obtain the DMM reading value by using the following formula:
data = DMM reading value (V) / Standard resistance value ()
9. Set the calculated data by XD data.
10. Read and check the measurement data.
11. Move to the next step
When changing the current calibration range: XNXT
When moving to voltage source/voltage limit calibration:
XNXT XVS, XVLH, XVLL