6253 DC Voltage Current Source/Monitor Operation Manual
8.2 Source and Measurement Functions
8-9
8.2 Source and Measurement Functions
DC source and measurement: Source and measurement of DC voltage and current
Pulse source and measurement:Source and measurement of pulse voltage and current
(However, measurement auto range in pulse source is impossible.)
DC sweep source and measurement:
Source and measurement by linear, multi-slope linear, random, log and fixed
levels
Pulse sweep source and measurement:
Source and measurement by linear, multi-slope linear, random, log and fixed
levels (However, measurement auto range in pulse source is impossible.)
Source value monitor: Monitoring of DC voltage current source values (Separately from the
measurement function)
Integration time: 10 types available: 5 s, 10 s, 100 s, 500 s, 1 ms, 10 ms, 1 PLC, 2 PLC,
200ms and arbitrary value (variable integration)
(PLC: Power Line Cycle 50 Hz: 20 ms, 60 Hz: 16.66 ms)
Variable integration range: 100 s to 1000 ms (setting resolution: 100 s)
Sweep reverse mode: ON (reverse) / OFF (single)
Sweep repeat count: 1 to 1000 times or infinite
Maximum number of sweep steps:
20000 steps
Maximum random sweep memory:
20000 data
Sampling count: 1 to 20000 times
(Plural samplings done by single trigger)
(Enabled only when setting to HOLD in the DC or pulse mode)
Measurement data memory: 20000 data
Measurement auto range: Available only in VSIM or ISVM
Measurement function link mode:
Links the measurement function to the source function.
VSIM or ISVM, ON/OFF available
Limit: The HI and LO limit values can be set individually.
(However, setting the current limit values with the same polarity is not
allowed.)
Calculation function: NULL calculation
Comparator calculation (HI, GO, or LO)
Scaling calculation
MAX, MIN, AVE, TOTAL calculations
Trigger style: Auto trigger, External trigger
Output terminal: Front Safety socket
Rear 5-pin rectangular
HI OUTPUT?HI SENSE?LO OUTPUT?LO SENSE?
DRIVING GUARD