6253 DC Voltage Current Source/Monitor Operation Manual
8.1 Source and Measurement
8-8
Effective CMRR: With unbalanced impedance of 1 k
At DC and AC 50/60 Hz 0.08%
NMRR: At AC 50/60 Hz 0.08 %
Integration time
5 s to 10 ms 1 PLC to 200 ms
Current source/
voltage measurement
60 dB 120 dB
Integration time
5 s to 10 ms 1 PLC to 200 ms
Voltage measurement/
current measurement
0 dB 60 dB